EnglishAnglų
phone
Telefonas:0312-3189593
email
Bal . 08, 2025 15:12 Grįžti į sąrašą

oil flash point tester



1. Technical parameters
Display: Color LCD display
Operation mode: Touch screen
Measuring range: 40~300℃
Temperature detection: Platinum resistance
Accuracy: ≥110℃±2℃ ≤110℃±1℃
Ignition mode: Electric ignition
Information storage: Can store 1000 measurement results
Cooling method: Forced air cooling
Printer: Thermal, Chinese characters, 40 lines
Self-test function: Test arm, igniter, printer, etc.
Repeatability: Meets ASTM D93 GB/T 261 standards
Power supply: AC 220V±11V, 50Hz±2.5HZ
Power: ≤350VA
Ambient temperature: 10℃~35℃
Ambient humidity: ≤85%

2. Working principle
The instrument is in accordance with ASTM D93, GB/T261-2008 method specified heating curve, automatic ignition when the temperature is close to the flash point value, when the flash point appears, the instrument automatically locks the display and prints the result, and automatically cools the heater.


Jei jus domina mūsų produktai, galite palikti savo informaciją čia ir mes su jumis susisieksime artimiausiu metu.