英語
電話:0312-3189593

PUSH electricalPS-2000A insulating oil dielectric loss and volume resistivity tester

PS-2000A Insulating oil dielectric loss tester /Oil tan delta tester is new generation of fully automatic portable insulating oil dielectric loss and resistivity meter designed according to ASTM D924 and GB/5654.7-inch color touch screen It is used for measuring dielectric dissipation factor and DC resistivity of insulating oil and other insulating liquids. It is easy to operate with reliable and accurate test data. Model No.: PS-2000A
PDFをダウンロード
詳細
タグ
製品のセールスポイント紹介
  1. High automation, heating, measuring dielectric loss, and measuring resistivity can be completed at a time.
  2. The three electrode type structure with an GB/T5654-2007 standard, inter electrode spacing 2mm, can eliminate stray capacitance and leakage effect on the dielectric loss test results.
  3. The instrument uses medium frequency induction heating, PID temperature control algorithm. This heating mode has the advantages of non-contact, oil cup and heating body, uniform heating speed, convenient control, the temperature control in temperature within the preset range error.
  4. Using advanced DSP and FFT technology to ensure the stability, accuracy and reliability of the data
  5.  Internal standard capacitor for the SF6 charging three pole capacitor, dielectric loss and capacitance of the capacitor is not affected by ambient temperature, humidity, etc., so that the accuracy of the instrument after a long time to use is still guaranteed.
  6.  Large screen color touch screen, man-machine dialogue convenient, concise operation, clear.
  7.  With the lid off the high voltage, high voltage electrode cup short reminder, eliminate safety hazards, to ensure the normal operation of the safety of operating personnel and equipment.
  8.  With real-time clock, test date, time can be saved with the test results, display, print; equipment can display the ambient temperature, real-time detection of the test environment.
  9.  Automatic storage measurement data, can store 100 sets of measurement data.
  10.  Correction function of empty electrode cup. The capacitance and dielectric loss factor of the empty electrode cup are measured to judge the cleaning and assembling condition of the empty electrode cup. The calibration data are saved automatically to facilitate the accurate calculation of relative capacitance and DC resistivity.
  11. Automatic emptying cup without disassembly, convenient and quick.
製品パラメータ

 

パラメータ

索引

パラメータ

索引

測定
範囲

キャパシタンス

5pF~200pF

解像力

キャパシタンス

0.01pF

誘電損失

0.00001~100

誘電損失

10-5

抵抗率

2.5MΩm~20TΩm

抵抗率

0.001M

測定
正確さ

キャパシタンス

0.5%+1PF

制御精度

±0.5℃

誘電損失

±(1%読み取り+0.0001)

温度範囲

0~125℃

抵抗率

±10%読み取り

交流電圧

交流0〜2200V

直流電圧

DC0〜600V

周囲温度

0~40℃

周囲の湿度

80%RH未満

動作電源

AC220V±10%(50±1)Hz

サイズ

420mm×380mm×420mm

100W

重さ

21Kg(フリーカップ)

 

ビデオ

 

弊社の製品にご興味がございましたら、こちらに情報を残していただければ、すぐにご連絡させていただきます。


ここにメッセージを書いて送信してください
関連している ニュース
  • Using Distillation Range Testers in the Food and Beverage Industry
    Using Distillation Range Testers in the Food and Beverage Industry
    The food and beverage industry relies on distillation to refine essential ingredients, from flavor extracts to alcoholic beverages and edible oils.
    詳細
  • The Impact of IoT on Distillation Range Tester Performance
    The Impact of IoT on Distillation Range Tester Performance
    The Internet of Things (IoT) is transforming industries worldwide, and the field of distillation range testing is no exception.
    詳細
  • The Best Distillation Range Testers for Extreme Conditions
    The Best Distillation Range Testers for Extreme Conditions
    In the world of chemical engineering and laboratory testing, precision and reliability are paramount.
    詳細

弊社の製品にご興味がございましたら、こちらに情報を残していただければ、すぐにご連絡させていただきます。